Philips M3015B Microstream CO2 Module REF 865377 #C08

Philips M3015B Microstream CO2 Module REF 865377 #C08

Philips M3015B CO8 module template

Performed limited functional verification using Philips IntelliVue MP30 and Pronk SimSlim patient simulator. Module recognized by monitor without reported technical faults. Microstream sampling pump activated upon FilterLine connection. CO2 response and capnogram observed using exhaled breath. Temperature input responded to applied temperature change. Both invasive pressure channels zeroed and tested at 200 mmHg using simulator; readings within manufacturer-specified ±5 mmHg tolerance.

CO2 gas accuracy, sampling flow, and internal leakage were not verified. No CO2 calibration performed.

Condition: Used. Clean physical condition overall. One dent at lower back of module. See pic.