 ***As-Is. For parts or repair. Head position sensor/indicator circuit FAILS, leaving the shutters stuck in the open position once they have first been triggered by the optical sensor after startup. The manual right and left occlude shutters function fine.*** 

Includes-
Control Panel and corresponding connecting cord. 

Included slides-
Slide 1: SMPS-B2 (13.26.33 Rev.B)
Slide 2: SMPS-30-1 (13.26.01 Rev.B)
Slide 3: SMPS-30-2 (13.26.02 Rev.B)
Slide 4: SMPS-30-3 (13.26.03 Rev.B)
Slide 5: SAS-2 (13.26.57 Rev.C)
Slide 6: SPS-3 (13.26.41 Rev.B)
Slide 7: SPCD-1 (13.26.34 Rev.E)
Slide 8: SA-BRL-4 (13.26.28 Rev.C)

DOES NOT INCLUDE- intermediate lenses.

Power on: OK
Tilt mechanism: OK
Lenses: OK
Light source condition indicator: OK (steady green after 1 min)
Peripheral vision screening left and right: OK
Test distance indicator for NEAR and Far (INTERMEDIATE not tested; no lens to test with): OK
Day/night selector: OK
Slide advance/reverse and position indicator: OK
Slide reset: OK
Occlude right and left: OK
Head positions indicator and override: ***FAIL*** 

Missing rubber hold down ring for slides. Headrest has discoloration.