DNI Nevada/ Fluke Biomedical MedTester 5000C

DNI Nevada/ Fluke Biomedical MedTester 5000C

DNI Nevada/ Fluke Biomedical MedTester 5000C
Version 5.12

***Calibration overdue***
Last Calibration Expired on 13-Mar-2026
***No cables or accessories included***

POST- Pass
Buttons- Function
Display- Good
Clock- Good
10 Lead output functional test- Pass
Mains receptacle test outlet- Functions
Kelvin cable connectors- Function
No other testing done.

Electrical Safety Test of MedTester 5000 (using Pronk Safe-T Sim): 
Ground resistance= 36 mOhms, 
Leakage Current= 112 microAmps,
PASS

Condition- Used. Very Good.